Symmetry Surgical is now part of Aspen Surgical.

Bovie® Surgi-Center | PROâ„¢

A2350

Bovie® Surgi-Center Pro High Frequency Electrosurgical Generator

The Surgi-Center|PRO (A2350) is a 200 watt high power, operating room generator, by Bovie. The unit incorporates the power and safety of a premium OR generator at a significant savings to the medical facility. The system delivers 12 monopolar and bipolar RF energy modalities to provide a high level of surgical precision. Engineered with the safety of the patient in mind, the Bovie Surgi-Center|PRO includes technology such as BovieNEM™ return split pad contact quality monitoring, BovieFDFS™ tissue sensing technology, self-test circuits, as well as specialized modalities for laparoscopic and delicate procedures. The Bovie Surgi-Center|PRO is supported by Bovie’s unprecedented 4-year standard warranty.

Request a Quote

MONOPOLAR WATTS VPEAK MAX VOLTAGE CREST FACTOR (TYPICAL) FREQUENCY (KHZ)
Cut I 200 1000V 1.7 ± 20% 490 kHz ± 4.9 kHz
Cut II  200 750V 1.7 ± 20% 490 kHz ± 4.9 kHz
Blend (1) 200 1320V 1.8 ± 20% 490 kHz ± 4.9 kHz
Blend (2) 200 1475V 2.0 ± 20% 490 kHz ± 4.9 kHz
Blend (3) 200 1650V 2.2 ± 20% 490 kHz ± 4.9 kHz
Blend (4) 200 1870V 2.4 ± 20% 490 kHz ± 4.9 kHz
Pinpoint Coagulation 120 1800V 3.1 ± 20% 490 kHz ± 4.9 kHz
Spray Coagulation 120 4000V 6.0 ± 20% 350 to 450 kHz
Gentle Coagulation 120 450V 1.6 ± 20% 490 kHz ± 4.9 kHz
BIPOLAR WATTS VPEAK MAX VOLTAGE CREST FACTOR (TYPICAL) FREQUENCY (KHZ)
Macro Bipolar 80 600V 1.5 ± 20% 490 kHz ± 4.9 kHz
Micro Bipolar 80 500V 1.5 ± 20% 490 kHz ± 4.9 kHz
Standard Bipolar 50 250V 1.5 ± 20% 490 kHz ± 4.9 kHz
Cooling System Passive Cooling
Technology Isolated RF Output
Audible/Visible Output Meter Yes
Pre-set Feature Yes
Monopolar HP Ports 2 Port(s)
Connector Ports/Power Switch White
Warranty 4 years
Contact Monitoring Systems Bovie NEMâ„¢
Width 37.5 cm (14.75 in)
Contact Monitoring Systems Range/Trip Point 10 ohms or less over entire operating range (10C° to 40°C)
Dual Activation Tone Alert Yes
Depth 46 cm (18.1 in)
Weight < 9.07 kg (<20 lbs)
Height 16.5 cm (6.5 in)

SPECIALIZED LAPAROSCOPIC MODE

Bovie’s Cut II is tailored specifically for Laparoscopic procedures. The mode offers a precisely managed cutting energy that limits output thereby minimizing the potential risk of capacitive coupling and alternate site burns.

DUAL MONOPOLAR SPRAY ACTIVATION

The SurgiCenter|PRO offers simultaneous activation in spray with audible activation tones. The simultaneous delivery produces equal and even power outputs in each accessory, for a safer delivery of dual monopolar energy.

ADJUSTABLE POWER DURING ACTIVATION

The power can be adjusted up or down in variable steps while performing a procedure, to increase procedural efficiency and provide the appropriate amount of power to the surgical site.

SURGICAL MEMORY TECHNOLOGY

The SurgiCenter | PRO recalls the last used power and mode. The surgeon can store 10 user-defined RF presets for easy recall.

PATIENT SAFETY FEATURES

  • This return electrode contact quality monitoring system is designed to minimize the risk of patient burns at the return electrode site. This feature works in conjunction with disposable split return electrodes.
  • Bovie FDFS™ – The tissue sensing technology measure tissue 5 million times a second and adjusts to varying impedances 5000 times a second for accurate energy delivery.
  • Automatic system self-checks upon powering up the unit.

ENERGY SETTINGS

 
 

CUT

Cut I

A general cutting with a standard RF waveform used in open and endoscopic surgical procedures, Cut I is for use in a wide range of tissue with minimal hemostasis.

Cut II

Tailored specifically for Laparoscopic procedures, Cut II offers a precisely managed cutting energy that limits output for minimizing the potential risk of capacitive coupling and alternate site burns due to weakened insulation.

Blend (Level 1, 2, 3 & 4)

Blended cutting offers varying levels of hemostasis during a cut procedure providing increased control of bleeding at the surgical application site.

  • Level 1: 75% Duty Cycle
  • Level 2: 62.5% Duty Cycle
  • Level 3: 50% Duty Cycle
  • Level 4: 37.5% Duty Cycle
 
 
Cut
 

COAGULATION

Pinpoint

A general coagulation mode delivering rapid desiccation effects with minimal tissue damage.

Spray

Recommended for hemostasis in highly vascular or large surface areas and for non-contact tissue coagulation to control superficial bleeding. Spray mode permits dual, simultaneous activation of two monopolar accessories with audible dual activation tone alert.

Gentle

Gentle mode offers more concentrated coagulation than in other modes. When coagulation is necessary in short amounts of time, reduced electrode carbonization is provided.

 
 
Coag
 

BIPOLAR

Macro

Intense bipolar output allowing for aggressive coagulation, with cut-like effects over a wide range of tissue.

Micro

Fine bipolar coagulation for use in extremely gentle and precise surgical applications.

Standard

Bipolar coagulation which gently coagulates tissue for minimal tissue damage reducing carbonization and sticking. Maximum power is 50 watts.

 
 
Bipolar
 

RECEPTACLE PANEL

 
 
Monopolar 1

Monopolar 1 Handswitching/
Footswitching Receptacle

Accepts standard 3-pin handpieces or accepts footswitching adapter (A1210A) or cable (A1210D or A1210R). Connect handswitching accessories.

 

 
Monopolar 2

Monopolar 2 Handswitching Receptacle

Accepts standard 3-pin handpieces. Connect handswitching accessories.

 
Bipolar

Bipolar Receptacle

Accepts standard cables for bipolar handpieces. Connect bipolar accessories.

ACCESSORIES