Bovie® IDS-310

IDS-310

Bovie High Frequency Electrosurgical Generator

The Bovie IDS-310 Electrosurgical Generator was designed in conjunction with OR clinicians to provide a safe, user-friendly generator to meet all operating room procedure performance demands. This innovative energy-based device offers 13 modes of surgical energy delivery. Special modalities include the advanced Bovie Bipolar, Auto Bipolar options and special Laparoscopic modes. These modes were included to assist surgeons in tailoring all electrosurgery needs for their variety of surgical applications and user preferences.

Request a Quote

MONOPOLAR WATTS VPEAK MAX VOLTAGE CREST FACTOR (TYPICAL) FREQUENCY (KHZ)
Cut I 300 1000V 1.7 ± 20% 490 kHz ± 4.9 kHz
Cut II 300 750V 1.7 ± 20% 490 kHz ± 4.9 kHz
Blend (1) 200 1320V 1.8 ± 20% 490 kHz ± 4.9 kHz
Blend (2) 200 1475V 2.0 ± 20% 490 kHz ± 4.9 kHz
Blend (3) 200 1650V 2.2 ± 20% 490 kHz ± 4.9 kHz
Blend (4) 200 1870V 2.4 ± 20% 490 kHz ± 4.9 kHz
Pinpoint Coagulation 120 1800V 3.1 ± 20% 490 kHz ± 4.9 kHz
Spray Coagulation 120 4000V 6.0 ± 20% 350 to 450 kHz
Gentle Coagulation 120 450V 1.6 ± 20% 490 kHz ± 4.9 kHz
BIPOLAR WATTS VPEAK MAX VOLTAGE CREST FACTOR (TYPICAL) FREQUENCY (KHZ)
Macro Bipolar 150 600V 1.5 ± 20% 490 kHz ± 4.9 kHz
Micro Bipolar 80 450V 1.5 ± 20% 490 kHz ± 4.9 kHz
Standard Bipolar 50 300V 1.5 ± 20% 490 kHz ± 4.9 kHz
Bovie Bipolar 225 250V 1.5 ± 20% 490 kHz ± 4.9 kHz
Auto Bipolar 80     490 kHz ± 4.9 kHz
Cooling System Passive Cooling
Technology Isolated RF Output
Audible/Visible Output Meter Yes
Pre-set Feature Yes
Monopolar HP Ports 2 Port(s)
Connector Ports/Power Switch Black
Warranty 4 years
Contact Monitoring Systems Bovie NEMâ„¢
Width 37.5 cm (14.75 in)
Contact Monitoring Systems Range/Trip Point 10 ohms or less over entire operating range (10C° to 40°C)
Dual Activation Tone Alert Yes
Depth 46 cm (18.1 in)
Weight < 9.07 kg (<20 lbs)
Height 16.5 cm (6.5 in)

BOVIE BIPOLAR (BBP)

Bovie Bipolar is a pulsing coagulation for quick and powerful coagulation of tissue. The pulsing performance can be tailored specifically to achieve the optimal surgical effect.

AUTO BIPOLAR INTUITIVE TECHNOLOGY

Auto Bipolar offers immediate and simplified tissue coagulation with minimal tissue damage. This feature automatically activates, coagulates and deactivates for precise tissue coagulation and eliminates the need for a footswitch. This feature is compatible with Bovie’s Micro, Macro and Standard Bipolar modes and can be set to surgical preferences for critical procedures.

SPECIALIZED LAPAROSCOPIC MODE

The IDS-310 recalls the last used power and mode. The unit’s user preset can recall up to 10 presets and up to 6 bipolar output preferences.

ADJUSTABLE POWER DURING ACTIVATION

During all activations, the power can be adjusted up or down in variable steps, giving surgeons more control at the surgical site.

PATIENT SAFETY FEATURES

BovieNEM™ – This return electrode contact quality monitoring system is designed to minimize patient burns at the return electrode site. Bovie FDFS™ – The tissue sensing technology measure tissue 5 million times a second and adjusts to varying impedances 5000 times a second for accurate energy delivery. A visual Ammeter indicates real time current (mA) flowing through the patient.

ENERGY SETTINGS

 
 

CUT

Cut I

A general cutting with a standard RF waveform used in open and endoscopic surgical procedure, Cut I is for use in a wide range of tissue with minimal hemostasis.

Cut II

Tailored specifically for Laparoscopic procedures, Cut II offers a precisely managed cutting energy that limits output for the minimizing of the potential risk of capacitive coupling and alternate site burns due to weakened insulation.

Blend (Level 1, 2, 3 & 4)

Blended cutting offers varying levels of hemostasis during a cut providing increased control of bleeding at the surgical application site.

  • Level 1: 75% Duty Cycle
  • Level 2: 62.5% Duty Cycle
  • Level 3: 50% Duty Cycle
  • Level 4: 37.5% Duty Cycle
 
 
Cut
 

COAGULATION

Pinpoint

A general coagulation mode delivering rapid desiccation effects with minimal tissue damage.

Spray

Recommended for hemostasis in highly vascular or large surface areas and for non-contact tissue coagulation to control superficial bleeding. Spray mode permits dual, simultaneous activation of two monopolar accessories with audible dual activation tone alert.

Gentle

Gentle mode offers more concentrated coagulation than in other modes. For instance, when coagulation is necessary in short amounts of time, reduced electrode carbonization is provided.

 
 
Coag
 

BIPOLAR

Standard

Bipolar coagulation which gently coagulates tissue for minimal tissue damage reducing carbonization and sticking.

Macro

Intense bipolar output allowing for aggressive coagulation, with cut-like effects over a wide range of tissue.

Micro

Fine bipolar coagulation for use in extremely gentle and precise surgical applications.

Bovie Bipolar

Exclusive to the Bovie IDS-310 and Bovie OR|PRO 300, Bovie Bipolar is a pulsing bipolar coagulation for quick and powerful coagulation of tissue. Pulsing performance can be tailored specifically to achieve the optimal surgical effect.

 
 
Bipolar
 

RECEPTACLE PANEL

 
 
Monopolar 1

Monopolar 1 Handswitching/ Footswitching Receptacle

Accepts standard 3-pin handpieces or accepts footswitching adapter (A1210A) or cable (A1210D or A1210R). Connect handswitching accessories.

 
 
Monopolar 2

Monopolar 2 Handswitching Receptacle

Accepts standard 3-pin handpieces. Connect handswitching accessories.

 
 
Bipolar

Bipolar Receptacle

Accepts standard cables for bipolar handpieces. Connect bipolar accessories.

ACCESSORIES